Medical Physics
Picket Fence Analyzer AI
Automated MLC leaf analysis with MLC type and treatment mode detection (static / VMAT) and ANSM compliance.
AI
Analysing the images…
No analysis yet.
Upload one or several Picket Fence images, or run the demo image, to see the results here.
How it works
- Upload one or several Picket Fence EPID images — ideally one per cardinal gantry angle for the ANSM monthly control.
- The MLC type and the test mode (static, dynamic, VMAT) are read from the DICOM header; pylinac finds every picket and measures the position error of each leaf pair.
- Each file gets two statuses: against your tolerance, and against the ANSM criteria of the decision of 28 February 2023 (max error ≤ 0.6 mm and at least 95 % of leaves passing).
- Optionally, an AI expert drafts an ANSM-oriented interpretation from the measured values — always to be validated by a medical physicist.
Documentation: pylinac Picket Fence.