CatPhan phantom analysis (504, 503, 600, 604): HU linearity, uniformity, spatial resolution.

AI
CBCT Analyzer AI — Kaptan Data Solutions
Analysing the CBCT series (about 10 s)…

No analysis yet.

Upload a ZIP of CatPhan slices, or run the CatPhan 504 demo, to see the results here.

How it works

  1. Scan a CatPhan phantom, export the axial slices as DICOM and put them in one ZIP.
  2. pylinac locates each module and measures HU linearity, geometry and slice thickness (CTP404), uniformity and noise power (CTP486), spatial resolution / MTF (CTP528) and low contrast (CTP515).
  3. Each value is compared with your reference control, as required by the ANSM decision of 28 February 2023 (controls 4C.1.1 to 4C.1.6). Without reference values the status is « incomplete ».
  4. Download cbct_baseline.json after your reference control and reload it at every following control.

Documentation: pylinac CBCT.