Medical Physics
CBCT Analyzer AI
CatPhan phantom analysis (504, 503, 600, 604): HU linearity, uniformity, spatial resolution.
AI
Analysing the CBCT series (about 10 s)…
No analysis yet.
Upload a ZIP of CatPhan slices, or run the CatPhan 504 demo, to see the results here.
How it works
- Scan a CatPhan phantom, export the axial slices as DICOM and put them in one ZIP.
- pylinac locates each module and measures HU linearity, geometry and slice thickness (CTP404), uniformity and noise power (CTP486), spatial resolution / MTF (CTP528) and low contrast (CTP515).
- Each value is compared with your reference control, as required by the ANSM decision of 28 February 2023 (controls 4C.1.1 to 4C.1.6). Without reference values the status is « incomplete ».
- Download
cbct_baseline.jsonafter your reference control and reload it at every following control.
Documentation: pylinac CBCT.